Implication of CT Table Sag on Geometrical Accuracy During Virtual Simulation

Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston, TX

Received 30 April 2007; accepted 5 July 2007.

Abstract 

Computed tomography (CT) scanners are used in hospitals worldwide for radiation oncology treatment simulation. It is critical that the process very accurately represents the patient positioning to be used during the administration of radiation therapy to minimize the dose delivery to normal tissue. Unfortunately, this is not always the case. One problem is that some degree of vertical displacement, or sag, occurs when the table is extended from its base when under a clinical weight load, a problem resulting from mechanical limitations of the CT table. In an effort to determine the extent of the problem, we measured and compared the degree of table sag for various CT scanner tables at our institution. A clinically representative weight load was placed on each table, and the amount of table sag was measured for varying degrees of table extension from its base. Results indicated that the amount of table sag varied from approximately 0.7 to 6.6 mm and that the amount of table sag varied not only between tables from different manufacturers but also between tables of the same model from the same manufacturer. Failure to recognize and prevent this problem could lead to incorrectly derived isocenter localization and subsequent patient positioning errors. Treatment site-specific and scanner-based laser offset correction should be implemented for each patient’s virtual simulation procedure. In addition, the amount of sag should be measured under a clinically representative weight load upon CT-simulator commissioning.

Key Words: Computed tomography, Table, Sag, Alignment

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0958-3947(07)00119-7

doi:10.1016/j.meddos.2007.07.001

Medical Dosimetry
Volume 32, Issue 4 , Pages 233-236, Winter 2007