Medical Dosimetry
Volume 35, Issue 1 , Pages 19-27, Spring 2010

Measurement and Evaluation of Inhomogeneity Corrections and Monitor Unit Verification for Treatment Planning

  • Rishik Saxena, M.S.
  • ,
  • Patrick Higgins, Ph.D.

      Affiliations

    • Corresponding Author InformationReprint requests to: Patrick Higgins, Ph.D., Therapeutic Radiology-Radiation Oncology, University of Minnesota, 420 Delaware Street S.E., MMC 494, Minneapolis, MN 55455-0110

Therapeutic Radiology-Radiation Oncology, University of Minnesota, Minneapolis, MN

Received 23 July 2008; accepted 2 January 2009. published online 23 February 2009.

Abstract 

Heterogeneous lung and bone phantoms have been constructed for the purpose of testing monitor unit calculations at or near interfaces for different planning systems. Data have been acquired for 2 linear accelerators: a Varian 2300cd (6 and 25 MV) and an Elekta Synergy (6, 10, and 18 MV). We have reviewed Pinnacle and the correction-based, pre-AAA version of Eclipse planning systems with the intent of exploring the limits of these systems with energy and field size. Data were acquired from 2 × 2 to 10 × 10 cm2 field sizes over the available range of energies. Our measurements confirm that Pinnacle predicts doses mostly to within ± 5%, even near lung-tissue interfaces over the full range of energies and field sizes tested. The Eclipse-modified Batho and equivalent TMR algorithms overpredicted doses by 10% or more in the lung and near the lung-tissue interfaces if the field size was less than 10 × 10 cm2 when the energy was 18 MV or higher. At lower energies, the field size had to be at least 6 × 6 cm2 for calculated doses to be within 10% of measurement. For bone-tissue interfaces, doses were generally underestimated by 5% to 10% or more by all calculation methods over the range of field sizes and energies reviewed. A second goal of this study was to review methods for hand-checking monitor units when heterogeneities are included. We evaluated the range of applicability of 2, one-dimensional (1D) inhomogeneity correction factors: the effective attenuation method and the TMR ratio method. The effective attenuation method for monitor unit checking was within ± 5% to as small as 6 × 6 cm2 fields for 6 to 10 MV, useable for 4 × 4 cm2 fields (within 7%) for 6 MV and close to ± 5% for 10 × 10 cm2 fields in the 18- to 25-MV range. The TMR ratio method was not as good, being within about ± 5% to 7% of measurements only for 6 × 6 to 10 × 10 cm2 fields at 6 MV and 10 × 10 cm2 fields at the higher energies. Both simple 1D correction methods performed almost as well as Pinnacle for the bone-soft tissue cases. We recommend that if direct measurement of dose for heterogeneous treatment plans is not practiced, then one of these simple cross checks be performed to assure patient safety.

Key Words: Heterogeneities, Monitor unit checking, Effective attenuation, TMR ratio

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0958-3947(09)00002-8

doi:10.1016/j.meddos.2009.01.002

Medical Dosimetry
Volume 35, Issue 1 , Pages 19-27, Spring 2010