Volume 35, Issue 4 , Pages 324-329, Winter 2010

Evaluation of a 3D Diode Array Dosimeter for Helical Tomotherapy Delivery QA

Moffitt Cancer Center, Division of Radiation Oncology, Tampa, FL

Received 26 May 2009; accepted 26 October 2009. published online 25 January 2010.

Abstract 

The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (<0.1%), dose rate dependence (0.4%), and absolute calibration accuracy (0.4% in the center of the phantom compared with the independently calibrated diode). Relative calibration of the arrays was verified by comparison with film and by rotating the detector 180°. The dosimeter response to rotational irradiation changed by no more than 0.2% when one of the detector boards was replaced by the homogeneous phantom material. The daily output correction factor can be derived from a Delta4 measurement in a uniform cylindrical field. The γ(3%, 3 mm) passing rate (absolute dose) was above 90% for all 9 evaluated clinical plans, and above 96% for all but one. The mean passing rate was 97 ± 2.7%. The plans varied in modulation factor, pitch, and calculation grid size. For best results, the phantom needs to be aligned carefully, preferably by megavoltage computed tomography imaging.

Key Words: Helical tomotherapy, Delivery quality assurance, Diode array dosimeter, Detector validation

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0958-3947(09)00126-5

doi:10.1016/j.meddos.2009.10.007

Medical Dosimetry
Volume 35, Issue 4 , Pages 324-329, Winter 2010